Skip to content

Discrepancy in SIMIND simulation noise levels for Ac-225 Kernel Fitting (Tutorial / PMB Paper) #13

@yhhuang98

Description

@yhhuang98

Hi there,

First of all, thank you so much for developing this toolbox. It is an incredibly valuable tool for the community!

I am currently trying to use the toolbox and SIMIND to perform Ac-225 440 keV SPECT PSF modeling. Following your methodology, I am simulating point source scanning at different source-detector distances and planning to fit it using the 1D operator

However, I have run into an issue at the very first stage regarding the raw Monte Carlo simulation outputs. When I run the simulation using the parameters provided in the tutorial, my raw projection images are significantly noisier than the clean reference files (.h00/.a00 files in the datalarger folder) provided in your tutorial repository.

Your simulations:
Image

My simulations:
Image

Some simulation Details:
Tutorial Version: SIMIND V7.0.3 (Intel Win)
My Version: SIMIND V8.0 (Intel Linux)
Photons: 2,000,000,000 (/26:2000)
Detector Hits: Both your and my .res logs at the same distance show roughly same # of hits.

Despite the .res logs showing statistically identical runs, directly visualizing my raw output file versus the tutorial's raw output file reveals a stark difference in the noise levels. Could you please offer us some advice on this?

Thank you again for your time and for this excellent toolbox!

Rick

Metadata

Metadata

Assignees

No one assigned

    Labels

    No labels
    No labels

    Type

    No type

    Projects

    No projects

    Milestone

    No milestone

    Relationships

    None yet

    Development

    No branches or pull requests

    Issue actions